Determination of As (III) and As (V) in natural waters using differential pulse voltammetry.

Author(s)
R. Ahmed Nuclear Chemistry Division, PINSTECH, P.O. Nilore, Islamabad, Pakistan
S. A. Qureshi Nuclear Chemistry Division, PINSTECH, P.O. Nilore, Islamabad, Pakistan
Abstract
A voltammetric method has been worked out for the analysis of arsenic. Arsenic is measured at hanging mercury drop electrode by differential pulse voltammetry in 1M HCI background electrolytes. Arsenic (III) peak appears at –0.40V vs. Ag/AgCl reference electrode. Detection limit of the method is 20mg/L of arsenic. In the second method Arsenic is measured using a background electrolyte of H2SO4 and HCI. For the analysis of total arsenic the as (V) is reduced to As (III) by the addition of ascorbic acid or sulfur dioxide or sodium bromide and hydrazine sulfate. Precision of the methods is quite good which is apparent from the good linearity curves. Recoveries of the added Arsenic from natural water samples were satisfactory. Stability studies of As (III) were also done to see any conversion to As (V). Many water samples collected from Rawalpindi-Islamabad area were analyzed and the results are reported and discussed. Results from the present study are compared with arsenic levels from different other countries.