No Detail Found!

Empirical relations for initial and volume recombination losses in Argon filled three electrodes ionization chamber.

Author(s)
Muhammad Munir Pakistan Institute of Engineering and Applied Sciences (PIEAS), Islamabd, Pakistan
Nasir Ahmad Pakistan Institute of Engineering and Applied Sciences (PIEAS), Islamabd, Pakistan
Abstract
Measurement of saturation current with accuracy is a prerequisite of any ionization chamber employed to measure the radiation doses. Recombination losses detract the current reaching saturation values in an ionization chamber and thus pose uncertainties in the dose rate measurements. To investigate these losses, the (I,V) characteristics of a locally designed and fabricated chamber filled with argon gas at various pressures ranging from 50-200 psi were studied with a Cs-137 gamma ray source at a dose rate 23 mGy/h in a reproducible geometry. It has been found that the initial recombination losses are larger than general recombination losses at this exposure. Empirical relations to estimate the fractional loss due to initial and general recombination losses have been developed.
Publication Details
Page(s) 13-16
DOI DOI not available
Published Journal: Science International, Volume: 17, Issue: 1, Year: 2005

Request the full text from PASTIC

0

Citations

0

Downloads

98

Views