A three electrodes ionization chamber in cylindrical geometry has been designed and developed. The useful length and volume of the chamber are 24.3 cm and 729 cm3, respectively. The (I,V) characteristics of the chamber filled with argon gas at various pressures ranging from 50-200 psi have been studied with a Cs-137 gamma rays. The operating voltage has been found to be 600 volt at 200 psi argon gas pressure. The results indicated that when positive ions were collected for the output electrical signal instead of electrons the saturation current has increased by 1%. An empirical relation between the saturation current and argon gas pressure for the designed chamber has been developed. The experimentally measured sensitivity of the chamber when the argon gas pressure was 200 psi (1.38 MPa) is 2.34 x 10-12 A/cm3 per rad/hr which is 0.85% less than the theoretically calculated value.