A precise method is described for the determination of boric acid and boron in metal borides by potentiometric titration of the mannitol/boric acid complex with a strong alkali(sodium hydroxide). Titration data are evaluated by a multiparametric curve fitting procedure on the basis of model functions for this type of titration. The initial boric acid concentration, or parameters like the conditional acidity constraint and the association number of the mannitol/boric acid complex can be determined. For a sample of titanium boride, the standard deviation was 0.15% on a boron content of 29.18%.