Abstract
Carbon content in solid carbon samples of activated charcoal and graphite has been determined on a CS-244 carbon sulfur analyzer. The instrumentutilizes solid state infra red detector. Different combustion accelerators including Lecocell, Lecocell-II tin, copper and iron metal chips were used. The results reveal that copper and iron fluxes produce accurate and reproducible results with standard deviations 1.3 for pure graphite and 0.39 (iron flux) and 0.17 (copper flux) for charcoal samples.
Publication Details
Page(s) 201-202
DOI DOI not available
Published Journal: Science International, Volume: 16, Issue: 3, Year: 2004