No Detail Found!

In Vitro Assessment of Immediate Implant Placement in Fresh Extraction Sockets versus Healed Sites Using a Polyurethane Mandibular Bone Model

Author(s)
Yeshwant Alias Shreyash Ulhas Tuenkar Department of Periodontics, KAHER's KLE V.K. Institute of Dental Sciences, Belagavi, Karnataka, India
Vishwannath Hiremath Department of Maxillofacial Surgery, Dr Hiremath Hospital, 79, CHBS Layout Nagarbhavi Main Road Vijayanagar Bangalore 560040, India
Pranay Yajurvedi Department of Oral and Maxillofacial Surgery, Pacific Dental College and Hospital, Udaipur, Rajasthan, India
Aditya Narayan Shukla Department of Oral and Maxillofacial Surgery, Babu Banarasi Das College of Dental Sciences, India
Rangoli Srivastava Department of Public Health Dentistry, Faculty of Dental Sciences, SGT University, Gurugram , India
Sameer Gupta Department of Oral and Maxillofacial Surgery, Inderprastha Dental College and Hospital, Sahibabad, Ghaziabad, Uttar Pradesh, 201010, India
Abstract
Background: Immediate implant placement in fresh extraction sockets is an established treatment option, but its initial biomechanical environment differs from that of delayed placement in healed sites. This controlled in vitro study compared stability and interface parameters between the two placement models under standardized artificial mandibular bone conditions. Methods: Forty polyurethane Sawbones blocks simulating type II/III mandibular bone were allocated to a fresh extraction socket model (Group A, n=20) or a healed osteotomy model (Group B, n=20). Identical SLA-surfaced titanium implants (4.0 × 10 mm) were inserted by a single operator. Primary outcomes were implant stability quotient (ISQ) and insertion torque value (ITV); secondary outcomes were removal torque value (RTV), bone-implant contact (BIC), contact area and gap volume assessed by micro-computed tomography. The artificial model evaluated mechanical stability only and did not simulate biological healing. Results: Healed site models demonstrated significantly higher stability than fresh socket models. Mean ISQ was 72.4±4.8 in Group B versus 64.2±5.6 in Group A (p<0.001). Mean ITV was 38.6±6.2 Ncm versus 28.4±5.8 Ncm (p<0.001) and mean RTV was 32.4±5.4 Ncm versus 24.8±4.9 Ncm (p=0.002). Mean BIC was 68.3±7.4% in healed sites and 52.6±8.2% in fresh socket models (p<0.001). Gap volume was substantially lower in healed sites (4.8±2.1 mm³) than in fresh socket models (12.4±3.2 mm³) (p<0.001). Conclusion: Within the limits of this in vitro polyurethane mandibular model, healed site placement showed greater primary stability and closer implant-bone adaptation than immediate placement in fresh socket models. Immediate placement may therefore require additional stabilization measures and cautious loading decisions when initial stability is borderline.
Publication Details
Page(s) 99-102
Published Journal: Journal of Pioneering Medical Sciences, Volume: 15, Issue: s1, Year: 2026

Request the full text from PASTIC

Keywords
Micro-CT, Resonance Frequency Analysis, Primary stability, Dental Implants, Immediate Placement, Extraction Socket, Healed Site
References
Brånemark P.I.,Period Year,April Year .2026 .Osseointegrated Implants in the Treatment of the Edentulous Jaw. Scandinavian Journal of Plastic and Reconstructive Surgery Supplement, 16 : 1-132.
Hämmerle C.H.,April C.H. .2026 .Consensus statements and recommended clinical procedures regarding the placement of implants in extraction sockets. International Journal of Oral and Maxillofacial Implants, 19(suppl) : 26-28.
Araujo M.G.,Lindhe J. .2005 .Dimensional ridge alterations following tooth extraction. An experimental study in the dog. ” Journal of Clinical Periodontology, 00642(2005) : 1600-051X.
Schulte W.,April W. .2026 .[The Tübingen immediate implant in clinical studies]. ” Deutsche Zahnärztliche Zeitschrift, 33(1978) : 348-359.
Schwartz-Arad D.,G. Chaushu. “ D. .1997 .The ways and wherefores of immediate placement of implants into fresh extraction sites: a literature review. ” Journal of Periodontology, 68(10) : 915-923.
Javed F. .2013 .Role of primary stability for successful osseointegration of dental implants: factors of influence and evaluation. Interventional Medicine and Applied Science, 4(2013) : 162-167.
Meredith N.,April N. .2026 .Assessment of implant stability as a prognostic determinant. International Journal of Prosthodontics, 11(1998) : 491-501.
Botticelli D. .2004 .Hard-tissue alterations following immediate implant placement in extraction sites. ” Journal of Clinical Periodontology, 00565(10) : 1600-051X.
Sennerby L.,Meredith N. .2008 .Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications. ” Periodontology, 00267 : 0757-66.
Trisi P. .2008 .Implant micromotion is related to peak insertion torque and bone density. ” Clinical Oral Implants Research, 01679(2009) : 0501-471.
Anitua E.,April E. .2026 .Novel technique for the treatment of the extraction socket: a pilot study. European Journal of Oral Implantology, 8(2015) : 75-82.
Bernhardt R. .2012 .Comparison of bone-implant contact and bone-implant volume between 2D histological sections and 3D SRμCT slices. European Cells and Materials, 23 : 70-81.
Lioubavina-Hack N. .2005 .Significance of primary stability for osseointegration of dental implants. ” Clinical Oral Implants Research, 01201(2006) : 0501-250.
Rodrigo D. .01820 .Diagnosis of implant stability and its impact on implant survival: a prospective case series study. ” Clinical Oral Implants Research, 21(2010) : 0501-261.
Esposito M.,April M. .2026 .Timing of implant placement after tooth extraction: immediate, immediate-delayed or delayed implants? A Cochrane systematic review. European Journal of Oral Implantology, 3(2010) : 189-205.
Chen S.T.,April S.T. .2026 .Immediate or early placement of implants following tooth extraction: review of biologic basis, clinical procedures and outcomes. International Journal of Oral and Maxillofacial Implants, 19(suppl) : 12-25.
Novaes A.B.,Jr A.B.,April A.B. .2026 .Immediate implants placed into infected sites: a histomorphometric study in dogs. International Journal of Oral and Maxillofacial Implants, 13(1998) : 422-427.
Chen S.T. .2007 .A prospective clinical study of non-submerged immediate implants: clinical outcomes and esthetic results. ” Clinical Oral Implants Research, 01388(2007) : 0501-562.
Nkenke E. .2003 .Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants. ” Clinical Oral Implants Research, 00937(2003) : 0501-609.
Sennerby L.,April L. .2026 .Resonance frequency analysis for implant stability measurements. A review.” Integrated Diagnostic Update. vol. 1, (2015) : 1-11.
Calvo-Guirado J.L. . .Peri-implant bone loss clinical and radiographic evaluation around rough neck implants: a 5-year study. ” Clinical Implant Dentistry and Related Research, 20(2018) : 3-8.
Raghavendra S.,April S. .2026 .Early wound healing around endosseous implants: a review of the literature. International Journal of Oral and Maxillofacial Implants, 20(2005) : 425-431.
Degidi M. .2010 .Primary stability determination by means of insertion torque and RFA in a sample of 4,135 implants. Clinical Implant Dentistry and Related Research, 00302(2012) : 8208-507.
Lang N.P. .2011 .A systematic review on survival and success rates of implants placed immediately into fresh extraction sockets after at least 1 year. Clinical Oral Implants Research, 02372(2012) : 0501-66.
Sanz M.,Cecchinato D. .2000 .Current concepts on immediate loading of implants. Periodontology, 66(2014) : 255-271.
Buser D. .2000 .Implant placement post extraction in esthetic single tooth sites: when immediate, when early. when late?” Periodontology, 73(2017) : 84-102.