Comparative analysis of the retention of locator attachments and ball attachments in implant supported mandibular overdentures

Author(s)
Omer Masood Alam Shad Departments of Prosthodontic, Institute of Dentistry, CMH Lahore Medical College, Pakistan
Marij Hameed Departments of Prosthodontic, Institute of Dentistry, CMH Lahore Medical College, Pakistan
Atiq-ur-Rehman Departments of Prosthodontic, Institute of Dentistry, CMH Lahore Medical College, Pakistan
Hammad Hassan Department od Science of Dental Materials, Azra Naheed Dental College, Superior University, Lahore, Pakistan
Muhammad Hassan University College of Medicine and Dentistry, University of Lahore, Pakistan
Mariyam Masood Faryal Dental College, Lahore, Punjab, Pakistan
Abstract
Objective: To compare the retention of ball attachments versus locator attachments in implant-supported mandibular overdentures during 5000 insertion and removal cycles. Methodology: This quasi - experimental study was conducted from January 2022 to October 2023. The sample size was 20 edentulous mandibular models and overdentures adequately seated with their respective housings. The overdentures were subjected to 5000 cycles of dislodging force using a universal testing machine, and the maximum loads needed to disengage the overdenture were measured for each cycle. The force values were then calculated. The data were analyzed using SPSS 22. Results: There was a significant difference in retentive forces between ball and locator types of attachments (p<0.001). The locator type had better retentive forces (52.45±4.97) than the ball type (32.61± 4.28). During the cycles, there was a decrease in retention in both types of attachments. Pearson's correlation showed a strong negative correlation between the number of pulls and retention in both types of attachments. The locator type showed a significantly lower retention loss than the ball-type attachment. Conclusion: The locator attachment mechanism was more favorable from a clinical standpoint. The drop in retention was attributable to both attachment systems; it was substantially lower in the case of the locator attach-ment system.
Publication Details
Page(s) 434-437
DOI DOI not available
Published Journal: Rawal Medical Journal, Volume: 50, Issue: 2, Year: 2025

Request the full text from PASTIC

Keywords
Denture retention, Prosthodontics., Denture bases, overden-ture
References
Hassan H,Rafique A,Andleeb S,Javaid M,Ahmad H,Khalid S H .2023 .Post-insertion complaints associated with complete denture prosthesis in a tertiary care hospital. Rawal Med J, 48 : 200-15.
Thayer ML,Ali R. .2022 .The dental demolition derby: bruxism and its impact-part 3: repair and reconstruction. Brit Dent J, 232 : 775-82.
Wakam R,Benoit A,Mawussi KB,Gorin C .2022 .Evaluation of retention, wear, and maintenance of attachment systems for single-or two-implant-retained mandibular overdentures: a systematic review. Materials, 15 : 1933-7.
Subin K,Koli DK,Jain V,Pruthi G,Nanda A .2021 .Comparison of ridge resorption and patient satisfaction in single implant-supported mandibular overdentures with conventional complete dentures: A randomised pilot study. J Oral Biol Cran Res, 11 : 71-7.
Aroso C,Silva AS,Ustrell R,Mendes JM,Braga AC,Berastegui E .2016 .Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro study. J Adv Prosthodont, 8 : 21-9.
Gonçalves F,Campestrini VLL,Rigo-Rodrigues MA,Zanardi PR .2020 .Effect of the attachment system on the biomechanical and clinical performance of overdentures: A systematic review. J Prosth Dent, 123 : 589-94.
Gibreel M,Lassila LV,Närhi TO,Perea-Lowery L,Vallittu PK .2021 .Midline denture base strains of glass fiber-reinforced single implant-supported overdentures. J Prosth Dent, 126 : 407-12.
Chavez SA,Locator R-TX Retention Inserts .2021 .Retention of Implant-Retained Overdentures: A Comparison of Locator. , : 1-72.
Marin DOM,Leite ARP,Oliveira Junior NMd,Paleari AG,Pero AC,Compagnoni MA .2018 .Retention force and wear characteristics of three attachment systems after dislodging cycles. Braz Dent J, 29 : 576-82.
Yabul A,Dayan C,Geckili O,Bilhan H,Tuncer N. .2018 .Evaluation of volumetric wear of abutments on the retention loss of ball attachment systems in implant‐retained overdentures: An in vitro study. Clin Impl Dent Relat Res, 20 : 778-84.
Choi ,Bae J-H,Jeong C-M Huh .2017 .Retention and wear behaviors of two implant overdenture stud-type attachments at different implant angulations. J Prosth Dent, 117 : 628-35.
ELsyad MA,Elhaddad AA,Khirallah AS .2018 .Retentive properties of O‐ring and locator attachments for implantretained maxillary overdentures: an in vitro study. J Prosthodont, 27 : 568-76.
Matthys C,Vervaeke S,Besseler J,Doornewaard R,Dierens M,De Bruyn H M .2019 .Five years follow‐up of mandibular 2‐implant overdentures on locator or ball abutments: Implant results, patient‐related outcome, and prosthetic aftercare. Clin Impl Dent Relat Res, 21 : 835-44.
Salehi R,Shayegh SS,Johnston WM,Hakimaneh SMR .2019 .Effects of interim plant distance and cyclic dislodgement on retention of LOCATOR and ball attachments: an in vitro study. J Prosth Dent, 122 : 550-6.
Hussein HAH,Badr AMI,Agamy EMTM,Mohamed GF .2019 .Evaluation of Implant Stability in Three Implants Supported Milled Bar Mandibular Overdenture Cases: A 1-Year Follow-up Clinical Trial. Ind J Public Health, 10 : 2415-9.
Arora S,Mittal S .2017 .Comparative Evaluation of Alteration in Retention Force values of Different Attachment systems for implant overdenture over various time intervals: An in Vitro Study. Int J Innov Res Sci, 4 : 71-6.
Rozov R,Trezubov V,Gerasimov A,Kopylov M,Azarin G. .2020 .Clinical analysis of the short-term and long-term results of the implant-supported Trefoil dental rehabilitation in Russia. Stomatologiia, 99 : 50-7.
Wichmann N,Kern M,Taylor T,Wille S,Passia N.,Retention N. .2020 .and wear of resin matrix attachments for implant overdentures. J Mech Behav Biomed Mater, 110 : 103901.
Sassi AN,Todaro C,Isola G,Bortolini I,Rodriguez y Baena R,Storelli S .2022 .. Mastication Wear of Two Low Profile Attachment Systems for Overdenture: An In Vitro Study. Bio Med Res Int, : 6469890.
Elhelbawy NG,ELsyad MA,Soliman TA,Mahrous A .2022 .Resilient Stud Versus Bar Attachments for Inclined Implants Supporting Mandibular Overdentures. An In Vitro Study of Loading and Dislodging Strains. Int J Oral Maxillofac Implants, 37 : 982-88.
Mirchandani B,Zhou T,Heboyan A,Yodmongkol S,Buranawat B .2021 .Biomechanical aspects of various attachments for implant overdentures: A review. Polymers, 13 : 3248-54.
Gray D,Patel J .2021 .Implant-supported overdentures: part 1. Br Dent J, 231 : 94-100.
Kortam S,Abdeen RM .2021 .Prosthetic complications and maintenance of implant retained maxillary and mandibular overdentures constructed with fully balanced and lingualized occlusal schemes: Randomized control clinical trial. Egypt Dent J, 67 : 2321-32.
Kleis WK,Kämmerer PW,Hartmann S,Al‐Nawas B,Wagner W. .2010 .A comparison of three different attachment systems for mandibular two‐implant overdentures: Oneyear report. Clin Implant Dent Relat Res, 12 : 209-18.