No Detail Found!

Development and validation of RP-HPLC method for determination of famotidine and its application in quality control of different pharmaceutical dosage forms.

Author(s)
Syed Saeed-Ul-Hassan University College of Pharmacy, University of the Punjab, Lahore, Pakistan
Saiqa Ishtiaq University College of Pharmacy, University of the Punjab, Lahore, Pakistan
Irshad Ahmad Department of Pharmacy, The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Muhammad Ayub Department of Pharmacy, The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Nayab Khalid Department of Pharmacy, The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Muhammad Munawar Hayat Faculty of Pharmacy, Bahauddin Zakariya University, Multan, Pakistan
Muhammad Tayyab Ansari Faculty of Pharmacy, Bahauddin Zakariya University, Multan, Pakistan
Muhammad Uzair Faculty of Pharmacy, Bahauddin Zakariya University, Multan, Pakistan
Abstract
A precise and fast novel high-performance liquid chromatography method was developed and validated for the quantitative determination of Famotidine (FMT) in commercially available pharmaceutical dosage forms. An Agilent 1200 Series High Performance Liquid Chromatography (HPLC) system having the column C18 (5µm particle size, 150×4.6 mm) was used in this study and detection (diode array detector) was made at 280 nm. The mobile phase was acetonitrile, distilled water, triethylamine and phosphoric acid (49.9:49.9:0.1:0.1, v/v), isocratic elution under ambient temperature at flow rate of 1.5 mL min-1 with injection volume 5µL. In this method, the retention times for FMT pure, tablets and suspension were 0.787 min, 0.789 min and 0.839 minutes respectively. The new method was validated by different validation parameters. The procedure provided a linear response over the concentration range of 0.1-1.0 mg mL-1 (r2 =0.998) and equation was y=3902.6+18.651. The mean % recovery for inter-day (96.56%) and intra-day (97.36%) assuring a good precision and accuracy was 96-98%. The method was found to be very rapid and the overall assay time was less than 2 minutes and the results obtained were accurate, precise and selective enough to allow the determination of FMT in the presence of certain excipients.
Publication Details
Page(s) 810-815
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 35, Issue: 3, Year: 2013

Request the full text from PASTIC

0

Citations

0

Downloads

91

Views