The objective of the present study was to develop a simple, rapid and sensitive reversed phase high performance liquid chromatographic (RP-HPLC) analytical method with UV detection system for the quantitative determination of metronidazole in human plasma. The chromatographic separation was performed by using C18 RP column (250mm X 4.6mm, 5µm) as stationary phase and 0.01M potassium dihydrogen phosphate buffered at pH 3.0 and acetonitrile (83:17, v/v) as mobile phase at flow rate of 1.0 ml/min. The UV detection was carried out at 320nm. The method was validated as per the US FDA guideline for bioanalytical method validation and was found to be selective without interferences from mobile phase components, impurities and biological matrix. The method found to be linear over the concentration range of 0.2812µg/ml to 18.0µg/ml (r2 = 0.9987) with adequate level of accuracy and precision. The samples were found to be stable under various recommended laboratory and storage conditions. Therefore, the method can be used with adequate level of confidence and assurance for bioavailability, bioequivalence and other pharmacokinetic studies of metronidazole in human.