Flow injection potentiometric determination of silver (I) in photographic film developing solution samples by polyvinylchloride membrane silver (I) selective electrode.

Author(s)
Cihan Topcu Erzincan University, Faculty of Arts and Sciences, Department of Chemistry, Yalnizbag, Erzincan, Turkey
Fatih Coldur Erzincan University, Faculty of Arts and Sciences, Department of Chemistry, Yalnizbag, Erzincan, Turkey
Muberra Andac Ondokuz Mayis University, Faculty of Science, Department of Chemistry, Kurupelit-Samsun, Turkey
Osman Cubuk Erzincan University, Faculty of Arts and Sciences, Department of Chemistry, Yalnizbag, Erzincan, Turkey
Gulsah Saydan Kanberoglu Department of Chemistry, Faculty of Science, Yuzuncu Yil University, Van, Turkey
Bulent Caglar Erzincan University, Faculty of Arts and Sciences, Department of Chemistry, Yalnizbag, Erzincan, Turkey
Ibrahim Isildak Yildiz Technical University, Faculty of Chemical and Metallurgical Engineering, Department of Bioengineering, Davutpasa Campus, Esenler, 34210 Istanbul, Turkey
Abstract
A polyvinylchloride (PVC) membrane silver (I) selective electrode was developed for the determination of silver (I) ions in photographic film developing solutions. The electrode was embedded into a home-made flow cell and successfully used as a potentiometric detector in a flow injection potentiometric (FIP) system. Under the optimized flow injection conditions, the electrode exhibited a non-Nernstian slope of 67.20±0.80 mV in the linear range from 1.0×10-5 to 1.0×10-1 mol/L (R2 =0.9961) and the lowest value for limit of detection (LOD) was calculated as 7.1×10-6 mol/L. The electrode displayed good sensitivity, selectivity, repeatable and rapid response towards silver (I) ions. Lifetime of the electrode was determined as at least two months in FIP system. The electrode was also successfully applied for the potentiometric determination of silver (I) ions in photographic film developing solution samples in FIP system. The results attained from FIP system were compared with the atomic absorption spectrophotometry (AAS) results by using student’s t-test at 95 % confidence level and relative deviation (RD) between the two methods was
Publication Details
Page(s) 882-888
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 38, Issue: 5, Year: 2016