No Detail Found!

Radiation Absorbed Dose for Cobalt-60 Gamma Source in Phantoms for Different Materials.

Author(s)
Muhammad Akram The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Hafiz Muhibb Ullah Zulkafal Shaukat Khanum Memorial Cancer Hospital & Research Centre (SKMCH&RC), Lahore, Pakistan
Saima Altaf The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Khalid Iqbal Shaukat Khanum Memorial Cancer Hospital & Research Centre (SKMCH&RC), Lahore, Pakistan
Muhammad Afzal Khan The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Saeed Ahmed Buzdar The Islamia University of Bahawalpur, Bahawalpur, Pakistan
Abstract
Current practices in radiation therapy required high doses of radiation to be delivered with increased accuracy. Treatment planning task is exercised till an optimum dose distribution is achieved. The present reported work was performed to compare the various aspects of the cobalt-60 radiation beam therapy with fixed source-surface distance 70cm incident normally. This study was conducted in May 2012 at the Department of Radiation Physics of MD Anderson Cancer Centre, University of Texas, Houston, United States. Radiation doses were calculated in a solid phantom as well as in water phantom at different square field sizes and depths. It was noted that the rate of absorbed dose increased with the increase in the field size and decreased with the increase in depths. The rate of absorbed dose was found to be directly proportional to the increase in the square field size and inversely proportional to the increase in depth. Moreover, the solid phantom demonstrated more absorbed doses as compared to the water phantom.
Publication Details
Page(s) 264-267
DOI DOI not available
Published Journal: Journal of Pakistan Medical Association, Volume: 68, Issue: 2, Year: 2018

Request the full text from PASTIC

Keywords
Cobalt-60, Phantoms, Gamma radiations, Ion chamber
References
FerlayJ,ShinHR,BrayF,FormanD,MathersC,ParkinDM,,2010,Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008,Int J Cancer,127,2893,917
JemalA,BrayF,CenterMM,FerlayJ,WardE,FormanD,,,,CA Cancer J Clin,201,69,90
RavichandranR,,2009,Has the time come for doing away with Cobalt-60 teletherapy for cancer treatments,J Med Phys,34,63,5
JoshiCP,DhanesarS,DarkoJ,KerrA,VidyasagarPB,SchreinerLJ,,2009,Practical and clinical considerations in Cobalt-60 tomotherapy,J Med Phys,34,137,40
WapstraAH,AudiG,ThibaultC,,2003,The Ame2003 atomic mass evaluation: (I). Evaluation of input data, adjustment procedures,Nuclear Physics A,729,129,336
HealyBJ,van der MerweD,ChristakiKE,MeghzifeneA,,2017,Medical Linear Accelerators: Competing Technologies for External Beam Radiotherapy,Clin Oncol,29,110,5
National Research Council. Radiation Source Use and Replacement: Abbreviated Version. National Academies Press; 2008 May 25
BabaMM,Mohib-,-HaqM,KhanMA,,2013,Dosimetric consistency of Co-60 teletherapy unit-a ten years study,Int J Health Sci (Qassim,7,15,21
FletcherGH,,1963,,in Management of Cancer. JAMA,183,17,22
ChoforN,HarderD,LooeHK,KapschRP,KollhoffR,WillbornK,,2009,Mapping radiation quality inside photon-irradiated absorbers by means of a twin-chamber method,Zeitschrift für Medizinische Physik,19,252,63
FerlayJ,ShinHR,BrayF,FormanD,MathersC,ParkinDM,,2010,Estimates of worldwide burden of cancer in 2008: GLOBOCAN 2008,Int J Cancer,127,2893,917
JemalA,BrayF,CenterMM,FerlayJ,WardE,FormanD,,,,CA Cancer J Clin,201,69,90
RavichandranR,,2009,Has the time come for doing away with Cobalt-60 teletherapy for cancer treatments,J Med Phys,34,63,5
JoshiCP,DhanesarS,DarkoJ,KerrA,VidyasagarPB,SchreinerLJ,,2009,Practical and clinical considerations in Cobalt-60 tomotherapy,J Med Phys,34,137,40
WapstraAH,AudiG,ThibaultC,,2003,The Ame2003 atomic mass evaluation: (I). Evaluation of input data, adjustment procedures,Nuclear Physics A,729,129,336
HealyBJ,van der MerweD,ChristakiKE,MeghzifeneA,,2017,Medical Linear Accelerators: Competing Technologies for External Beam Radiotherapy,Clin Oncol,29,110,5
National Research Council. Radiation Source Use and Replacement: Abbreviated Version. National Academies Press; 2008 May 25
BabaMM,Mohib-,-HaqM,KhanMA,,2013,Dosimetric consistency of Co-60 teletherapy unit-a ten years study,Int J Health Sci (Qassim,7,15,21
FletcherGH,,1963,,in Management of Cancer. JAMA,183,17,22
Chiu-Tsao ST, Chan MF. Photon beam dosimetry in the superficial buildup region using radiochromic EBT film stack. Med Physics 2009; 36: 2074-83.
Chofor N, Harder D, Looe HK, Kapsch RP, Kollhoff R, Willborn K, et al. Mapping radiation quality inside photon-irradiated absorbers by means of a twin-chamber method. Zeitschrift für Medizinische Physik 2009; 19: 252-63.
Hill R, Mo Z, Haque M, Baldock C. An evaluation of ionization chambers for the relative dosimetry of kilovoltage x ray beams. Med Physics 2009; 36: 3971-81
Schreiner LJ, Joshi CP, Darko J, Kerr A, Salomons G, Dhanesar S. The role of Cobalt-60 in modern radiation therapy: Dose delivery and image guidance. J Med Phys 2009; 34: 133-6.
Andreo P, Cunningham JR, Hohlfeld K, Svensson H. Absorbed dose determination in photon and electron beams. An international Code of Practice. 2nd ed. Vienna: International Atomic Energy Agency; 2000
Party IW, Thwaites DI, DuSautoy AR, Jordan T, McEwen MR, Nisbet A, et al. The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration. Phys Med Biol 2003; 48: 2929-70.
Musolino SV. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water; Technical Reports Series No. 398; 2000
Iqbal K, Isa M, Buzdar SA, Gifford KA, Afzal M. Treatment planning evaluation of sliding window and multiple static segments technique in intensity modulated radiotherapy. Rep Pract Oncol Radiother 2013; 18: 101-6.
Tailor RC, Chu C, Followill DS, Hanson WF. Equilibration of air temperature inside the thimble of a Farmer type ion chamber. Med Phys 1998; 25: 496-502.
Rogers DW, Ross CK. The role of humidity and other correction factors in the AAPM TG21 dosimetry protocol. Med Phys 1988; 15: 40-8.
Mijnheer BJ. Variations in response to radiation of a nylon walled ionization chamber induced by humidity changes. Med Phys 1985; 12: 625-6.
Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG 51 protocol for clinical reference dosimetry of high energy photon and electron beams. Med Phys 1999; 26: 1847-70.
Musolino SV. Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water; Technical Reports Series No. 398; 2000.
0

Citations

0

Downloads

55

Views