Efficiency calibration of germanium gamma ray detectors.

Author(s)
Mahmud Ahmad Pakistan Institute Nuclear Science Technology, Nilore, Islamabad, Pakistan
Muhammad Anwar Pakistan Institute Nuclear Science Technology, Nilore, Islamabad, Pakistan
Shaikh M. Saleem Pakistan Institute Nuclear Science Technology, Nilore, Islamabad, Pakistan
Khunab Gul Pakistan Institute Nuclear Science Technology, Nilore, Islamabad, Pakistan
Naeem A. Khan Pakistan Institute Nuclear Science Technology, Nilore, Islamabad, Pakistan
Abstract
The experimental measurements of absolute full energy gamma-ray peak detection efficiencies of a 105 cc Ge(Li) detector and a 61 cc intrinsic germanium detector for energy ranges 0. 05-11 MeV and 0. 05- 8. 6 MeV respectively using standard radioactive isotopes and thermal neutron capture reactions are described. A three crystal Ge-Nal spectrometer was operated in Compton suppression mode for these measurements. The spectrometer was operated in pair mode for the determination of relative double escape peak detection efficiencies of the detectors from 2-10 MeV gamma-ray energy rang9 Liking gamma-ray cascades following thermal neutron capture in gamma and 5'''5"'Cr. The detection efficiency curves are fitted with polynomials using least squares fitting technique.