Ionization chambers play an important role in monitoring gamma-ray activity. This paper gives the design of an ionization chamber which has been fabricated and tested in our laboratory. The effective volume of the chamber is 516 cm3, the total length is 32 cm and the outer diameter is 7. 3 cm. Commercially available Argon gas has been used as fill gas. The materials, except the connectors, available in the local market were used in the construction of the chamber. Two materials, teflon and gyrolite, have been tested as insulators and have been found to be satisfactory. The response of the chamber up to 600 R/hr due to 60Co gamma-rays is linear. The experimentally determined sensitivity is 1. 51 x 1012 A/cm3 per R/hr when the gas pressure is 10 atmosphere. The detector has also been tested using spent fuel of PARR and it has been observed that it is capable of monitoring the activity of a single spent fuel element.