No Detail Found!

X-ray fluorescence analysis of hafnium in hafnium-zirconium binary solution by a correction factor method of matrix effects.

Author(s)
Riaz Qadeer Pakistan Institute Nuclear Science Technology, Islamabad, Pakistan
Javed Hanif Pakistan Institute Nuclear Science Technology, Islamabad, Pakistan
M. Saleem Chemistry Department, Quaid-e- Azam University (QAU), Islamabad, Pakistan
M. Afzal Chemistry Department, Quaid-e- Azam University (QAU), Islamabad, Pakistan
Abstract
Hafnium content of a hafnium-zirconium binary solution was determined by wavelength dispersive X-ray fluorescence spectrometer, using alpha-correction factor method of matrix effect. The solutions were presented to the spectrometer in simple polyethylene bottles. The method of fitting the measured Hf L 3i intensity to a non linear function of composition is also examined and the comparison between the two is made.
Publication Details
Page(s) 261-263
DOI DOI not available
Published Journal: Science International, Volume: 3, Issue: 4, Year: 1991

Request the full text from PASTIC

0

Citations

0

Downloads

89

Views