Calibration and use of multiple strain-Gauge bridge configurations in transducer application.

Author(s)
Rehmatullah J. Malik Sind Agricultural Research Institute, Tandojam, Pakistan
M. Saleh Soomro Sind Agricultural Research Institute, Tandojam, Pakistan
Syed Gulzar Ali Shah Sind Agricultural Research Institute, Tandojam, Pakistan
M. Uris Khaskheli Sind Agricultural Research Institute, Tandojam, Pakistan
Abstract
For the simple strain-gauged cantilever beam calibration results revealed that the theoretical and experimental values of sensitivity (?v/W) were consistent to within ± 7%. In case of the shunts incorporated into the system, the experimental and theoretical ± 2%. values The of forces obtained theoretical and were correct to within experimental values of K (bridge constants) for different bridge arrangements were correct to within ± 5%.