Some aspects of the operating characteristics of Na I (TI) scintillation detectors for gamma rays.

Author(s)
M. Yar Khan Physics Department, Gomal University, Dera Ismail Khan, Pakistan
M. Ayub Physics Department, Gomal University, Dera Ismail Khan, Pakistan
Saadullah Jan Physics Department, Gomal University, Dera Ismail Khan, Pakistan
Q. Nawaz Physics Department, Gomal University, Dera Ismail Khan, Pakistan
Abstract
An acceptable operating range of the high voltage, on the photomultiplier tube of a Thallium activated Sodium Iodide Na 1 (TI) scintillation detector for gamma rays, has been selected where both the signal strength S and the signal to background (B) ratio S/B are a maximum. This has been found by measuring the figure of merit S/B as a function of the voltage. The high voltage value is set at about the middle of the selected operating range. A further check on the above selection of the operating voltage can be provided by obtaining the gamma ray spectrum of a source by operating the detector at various value of the high voltage and then determining the resolution of the detector for a photopeak. The operating voltage is set where the resolution for the photopeak seems best.