No Detail Found!

Enhancement of trace metal atomic absorption signals through the optimization of physical factors attendant on aspiration rates.

Author(s)
M. Jaffar Chemistry Department, Quaid-i-Azam University (QAU), Islamabad, Pakistan
M. Ashraf Chemistry Department, Quaid-i-Azam University (QAU), Islamabad, Pakistan
M. Tariq Chemistry Department, Quaid-i-Azam University (QAU), Islamabad, Pakistan
Abstract
An optimization scheme based on instrumental components controlling the aspiration rates of water sample solutions in the flame atomic absorption technique is proposed. The methodology adopted stems from the quantitative study of the effect of such factors as fuel/oxidant flow rates, burner-slit, length, radii of aspriation and sampling capillaries, aspiration temperature and addition of organic solvent to the aqueous samples. The method is applied to the estimation of silver, copper, chromium, iron, strontium, zinc, cadmium, lead and mercury concentrations in drinking waters without the use of a preconcentration step. An overall hundred-fold enhancement in the absorption signals is achieved at a fairly high S/N ratio. The method is quantitative with a precision of ± 1% at the detection level of the individual trace metals.
Publication Details
Page(s) 175-178
DOI DOI not available
Published Journal: Pakistan Journal of Scientific and Industrial Research, Volume: 29, Issue: 3, Year: 1986