Based on the acid-medium reduction of mercury, a flameless atomic absorption method of general utility is proposed for the determination of inorganic mercury in public utility waters. Practical evaluation is made of the parameters governing the expelled mercury absorbance in an optical cell fabricated from inexpensive materials. These include dimensions of the reduction vessel, effluent flow rate and the volume of the solution as related to the optimum sensitivity conditions. Surface and under-ground water samples, originating from various local water supply system, ranging in depth from 30-60 feet and feeding, on the average, a population of 10, 000 or more, were analyzed for mercury content. A lower detection limit of 100 ng/l mercury for the proposed method is established with an average precision of ± 2.5% corresponding 2.5% confidence level for replicate measurements. The method has the potential of application to mercury assay in food stuffs and beverages.