No Detail Found!

Device entrapment: a rare complication of percutaneous coronary intervention.

Author(s)
Abdur-Rehman Department of Cardiology, Karachi Institute of Heart Diseases (KIHD), Karachi, Pakistan
Javaid Arif Khan Department of Cardiology, Karachi Institute of Heart Diseases (KIHD), Karachi, Pakistan
Javaid A Sial Department of Cardiology, National Institute of Cardiovascular Diseases, Karachi, Pakistan
Zafar Haleem Baloch Department of Anatomy, Sindh Medical College, Karachi, Pakistan
Abstract
Device entrapment is a rare complication of percutaneous coronary intervention. It has hazardous potentials for the patient. Emergent cardiac surgery is the only option after failure of retrieval devices. We have described here a case of a 55 years old male. During percutaneous coronary intervention, the balloon inflated only partially at its ends and entrapped in the lesion along with the stent. Multiple attempts at inflation failed and the patient developed severe chest pain for few minutes. Surgery was contemplated but final attempt at inflation was successful and the patient stabilized. It was an emergent situation and could have led to fatal outcome, although this patient escaped narrowly from any fatal outcome.
Publication Details
Page(s) 620-622
DOI DOI not available
Published Journal: Journal of College of Physicians and Surgeons--Pakistan : JCPSP, Volume: 21, Issue: 10, Year: 2011

Request the full text from PASTIC

0

Citations

0

Downloads

103

Views