Percentage discrepancies assessment between measured and calculated behaviour of percent depth dose in external beam radiotherapy.

Author(s)
Muhammad Isaa Department of Radiation Oncology, Shaukat Khanum Cancer Hospital and Research Center, Lahore, Pakistan
Khalid Iqbal Department of Physics, The Islamia University, Bahawalpur, Pakistan Department of Radiation Oncology, Shaukat Khanum Cancer Hospital and Research Center, Lahore, Pakistan
Muhammad Jahanzeb Ashraf Department of Physics, The Islamia University, Bahawalpur, Pakistan
Muhammad Afzal Department of Physics, The Islamia University, Bahawalpur, Pakistan
Saeed Ahmad Buzdar Department of Physics, The Islamia University, Bahawalpur, Pakistan
Abstract
The aim of this study was to calculate percentage discrepancies (PD) of the measured and calculated percentage depth doses (PDDs) values. The 6 MV photon beam produced by the Varian linear accelerator 2100 C/D was used in this study. PDDs, tissue maximum ratios (TMR) and phantom scatter factor (Sp) were measured using the PTW 31006 ionisation chamber in water phantom. PD between PDD values of the measured and calculated was ranging between 0.30% and 2.38%. Percentage discrepancies were also found higher against 20 cm depth in water for (20×20) cm2 field size. These discrepancies should be taken into account, while delivering any medical dose in radiation therapy centers.