Correlation of blood pressure and the ratio of S1 to S2 as measured by esophageal stethoscope and wireless bluetooth transmission.

Author(s)
Kyoung Hoon Lim Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Young Duck Shin Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Sang Hi Park Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Jin Ho Bae Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Hong Jae Lee Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Seon Jung Kim Department of Anesthesiology and Pain Medicine, Chungbuk National University Hospital, Republic of Korea, Korea
Ji Yun Shin Department of Biomedical Engineering, Chungbuk National University Hospital, Republic of Korea, Korea
Young Jin Choi Department of Surgery, Eulji University Hospital, Republic of Korea, Korea
Abstract
Objective: Esophageal stethoscope has the advantage of being non-invasive, easily placed and capability to monitor the heart sound. This study was designed to determine whether the ratio of S1 to S2 analyzed by esophageal stethoscope and wireless Bluetooth transmission can be accurate indicator that express the correlation with blood pressure. Methods: Total 33 adult male and female without cardiac disorder and with normal heart rhythm were selected randomly as the subjects of this Study. Two microphones were used with one for acquisition of heart sound by connecting it to the esophageal stethoscope while the other was used to measure the background noise in the operating room. After having transmitted the heart sound measured with the esophageal stethoscope to the receiver by using bluetooth module, it was saved in PC and outputted, following removal of noise in the operating room and the respiratory sound. S1 and S2 were measured with computation of the ratio of S1 to S2. Correlations between the systolic blood pressure with each of the S1, S2 and ratio of S1 to S2 were examined by using correlation analysis. Results: The ratio of S1 to S2 displayed the highest correlation with the systolic blood pressure, with S1 and S2 also displaying positive correlation with the systolic blood pressure. Conclusion: As the result of analysis of the heart sound and the systolic blood pressure measured by using the esophageal stethoscope, the radio of S1 to S2 displayed greater correlation with the systolic blood pressure in comparison to the S1.