No Detail Found!

Counting statistics of Uranium determination by X-ray Fluorescence Spectrometry.

Author(s)
Riaz Qadeer Pakistan Atomic Energy Commission, Islamabad, Pakistan
Nasir Khalid Chemistry Division, Pakistan Institute of Nuclear Science and Technology, Post Office Nilore, Islamabad, Pakistan
Abstract
The procedure has been explained to evaluate the equipment performance by repeated measurements of energy dispersive X-ray fluorescence signal of 4.201 × 10−4mol L−1uranium solution in 3 mol L−1of HNO3. Standard statistical models were applied to the data. The Poisson distribution function [F(X)P] and Normal distribution function [F(X)N] were found to be 1.0197 and 1.0049 respectively, which are quite near to the ideal value of unity. Linearity range for uranium determination by XRF was found to be up to 1.47 × 10−2mol L−1of uranium, with regression coefficient “R2” of 0.998 under the optimized instrumental parameters.
Publication Details
Page(s) 44-49
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 36, Issue: 1, Year: 2014