No Detail Found!

Temperature regulation and standardization practices of clinical laboratories in Karachi.

Author(s)
Malik Muhammad Adil Shifa College of Medicine, Islamabad, Pakistan
Ali Yawar Alam Department of Community Health Sciences, Shifa College of Medicine, Islamabad, Pakistan
Abstract
Objective: To examine the temperature regulation and standardization practices of clinical laboratories in Karachi. Method: Forty five clinical laboratories in Karachi were examined for observing the standard protocols for running a lab with particular reference to temperature regulation. A questionnaire to the effect was filled. Results: Among the 45 labs included, the mean complete blood count performed per day was 52 ± 47. Only 5 (11%) labs had a temperature reader. Thirty (66.7%) labs had an air-conditioner installed, of which only 24 were found in working condition. Maintenance of instruments was carried out every 67.5 ± 30.6 days. The mean number of haemolyzed samples was 2.3 ± 1.7. Control was available in 24(53.3%) labs, which was used daily in only 10 labs. Quantity of blood was the same in all the tubes in only 33(73.3%) labs. Conclusion: Commercial laboratories should be properly registered and their quality standardized.
Publication Details
Page(s) 88-90
DOI DOI not available
Published Journal: Journal of Pakistan Medical Association, Volume: 55, Issue: 2, Year: 2005

Request the full text from PASTIC

0

Citations

0

Downloads

59

Views