No Detail Found!

Potentiometric determination of Potassium using an indigenously fabricated coated wire ion selective electrode.

Author(s)
Amber R. Solangi National Centre of Excellence in Analytical Chemistry University of Sindh, Jamshoro, Pakistan
A. W. K. Khanzada National Centre of Excellence in Analytical Chemistry University of Sindh, Jamshoro, Pakistan
Abstract
A potentiometric sensor system based on coated wire has been developed indigenously for selectivity towards potassium ion. The electrode has been prepared by coating a silver wire conductor with PVC based film. The preferred system with an antibiotic Valinomycin in membrane shows a linear response towards potassium ions over a concentration range of lx10-1 M to lx 10-5M of KCl solution. A saturated calomel electrode was used as a reference electrode with LiCl3Ac as an internal filling solution. To avoid interferences, KCl was not used. The electrode was also tried for the determination of potassium ion in KCl tablets NEO-K, supplied by Zaffa, Pvt. Ltd. Pakistan. Reported value of potassium in tablets was 260mg/tablet and result found with coated wire electrode was 265 mg / tablet.
Publication Details
Page(s) 88-90
DOI DOI not available
Published Journal: Pakistan Journal of Analytical Chemistry, Volume: 5, Issue: 2, Year: 2004

Request the full text from PASTIC

0

Citations

0

Downloads

109

Views