Clinical utility and accuracy of a blood glucose meter for the detection of neonatal hypoglycemia.

Author(s)
Muhammad Haroon Hamid Department of Paediatric Medicine, The Children’s Hospital and the ICH, Lahore, Pakistan
Akmal Laeeq Chishti Department of Paediatric Medicine, The Children’s Hospital and the ICH, Lahore, Pakistan
Sajid Maqbool Department of Paediatric Medicine, The Children’s Hospital and the ICH, Lahore, Pakistan
Abstract
To determine the accuracy and reliability of a glucometer in comparison to hexokinase method in detecting neonatal hypoglycemia. A comparative analytical study conducted at Neonatal Unit of the Children’s Hospital, Lahore, from August 2001 to February 2002. All neonates presenting with known risk factors or suggestive clinical features were screened for hypoglycemia by using capillary blood on Accutrend alpha glucometer. Simultaneously the venus blood glucose values were done on Hitachi 902 autoanalyser by hexokinase method. A level of 40 mg/dl or less was taken as neonatal hypoglycemia. A Total of 292-paired samples were taken from 223 neonates. Hypoglycemia was deteted in 112 samples (38.4%). Correlation of glucometer values with laboratory values of blood glucose levels was excellent throughout the range with coefficient of correlation ® of 0.976 (p-value
Publication Details
Page(s) 225-228
DOI DOI not available
Published Journal: Journal of College of Physicians and Surgeons--Pakistan : JCPSP, Volume: 14, Issue: 4, Year: 2004