Measurement of oxygen dissolved in water is a key parameter in water testing from industrial and environmental perspective. The classical Winkler method has been in practice since its invention, but the shortcomings such as loss of iodine, instability of endpoint and use of secondary standard thiosulfate makes it a time consuming process requiring highly skilled analysts to perform. In the present work, the procedure for estimation of DO has been modified from the iodometric to the redox titration wherein alkaline MnSO4 is used to fix DO; the precipitates formed as a result of fixation are dissolved by Fe+2 (Mohr’s Salt) solution used in excess. The amount of Fe+2 left un-reacted is determined by KMnO4. Each ml, of Fe+2 consumed by Mn+3 is equivalent to 1 ppm of DO.