Efficacy of esophageal detector device in verification of endotracheal tube placement.

Author(s)
Noor Hussain Fauji Foundation Hospital, Rawalpindi, Pakistan
Asadullah Jafri Combined Military Hospital (CMH), Rawalpindi, Pakistan
Rashad Siddiqi Combined Military Hospital (CMH), Rawalpindi, Pakistan
Abstract
To assess accuracy of esophageal detector device (EED) for detection of endotracheal tube placement and to compare its performance with that of capnography. A prospective study in which 400 patients were divided into two groups. In group-I the patients were intubated as per routine, then the EDD and capnograph were again used to check both the tubes. The result showed 100% sensitivity, 100% specificity and 100% predictive value for EDD. The EDD was found to be very effective device in differentiating esophageal from tracheal intubation.