Thermoplastic masks are widely used as an integral part of patient positioning and immobilization devices but their effect on dose distribution has not yet been quantified. Some manufacturers claim that these materials have very little effect on beam attenuation and dose buildup. However, no published data is available in the literature. To study the effect of these thermo-plastics on the dose distribution, three most commonly used thermoplastic sheets were selected for evaluation of dosimetric data and beam characteristics. Effects on central axis depth dose data, dose distribution profiles, surface dose and beam attenuation are presented in this article. No significant effect was found on the relative dose distribution pattern in terms of percentage depth doses and beam uniformity parameters, however, there was sign f cant amount of attenuation caused by the presence of the base-plate the head support cushions used with these masks during the treatment. In vivo dosimetry performed for selected number of patients also revealed that there was no change in the skin dose of the patient with or without the use of the thermoplastic masks.