No Detail Found!

Pulmonary artery catheter knotting in a coronary artery bypass surgery patient.

Author(s)
Mohammad Irfan Akhtar Department of Anaesthesia, The Aga Khan University, Karachi, Pakistan
Mohammad Hamid Department of Anaesthesia, The Aga Khan University, Karachi, Pakistan
Fazal Hameed Khan Department of Anaesthesia, The Aga Khan University, Karachi, Pakistan
Hamid Iqil Naqvi Department of Anaesthesia, The Aga Khan University, Karachi, Pakistan
Amar Gangwani Department of Anaesthesia, The Aga Khan University, Karachi, Pakistan
Abstract
A fifty seven years old female patient came to operating room for coronary artery bypass grafting (CABG) surgery. After induction, invasive monitoring lines were placed. Pulmonary artery catheter (PAC) floated after three attempts and it was wedged at 60cm. Intra operative course was smooth and patient transferred to cardiac intensive care unit (CICU). First chest X-ray revealed PAC knotting in the right ventricle. Vascular surgeon was involved and he removed it through right internal jugular vein under fluoroscopic guidance. Pulmonary arterial catheterization is an invasive procedure. Knotting usually occurs due to excessive advancement of the pulmonary artery catheter beyond the normally expected distance. The removal of a catheter should never be forced when resistance is encountered. PA catheter knotting is a rare complication but it should be suspected whenever there is excessive length of catheter required to reach pulmonary artery.
Publication Details
Page(s) 519-521
DOI DOI not available
Published Journal: Journal of Pakistan Medical Association, Volume: 57, Issue: 10, Year: 2007

Request the full text from PASTIC

0

Citations

0

Downloads

54

Views