Influence of polyether ether ketone coping crown on the adaptation of implant abutment

Author(s)
Xuesheng Li Department of Prosthodontics, Hainan Stomatological Hospital, Haikou, China
Shuai Liu Department of Prosthodontics, Shenzhen Jiahong Dental Co, Ltd, Shenzhen, China
Yishi Tang Department of Prosthodontics, Hainan Stomatological Hospital, Haikou, China
Yunshan Zhao Department of Prosthodontics, Hainan Stomatological Hospital, Haikou, China
Yingbin Su Department of Stomatology, Second Naval Hospital, China
Abstract
Objective: To evaluate the influence of polyether ether ketone coping crown on the adaptation of implant abutment. Method: The vitro study was conducted at the department of Prosthodontics, Hainan Stomatological Hospital, China, from October 2021 to March 2022, and comprised patients undergoing implant surgery on first molar. Patients were divided into two groups, with group A patients receiving polyether ether ketone coping crowns, and group B receiving zirconia crowns. Replica technique was used to replicate the gap between the crowns and the abutments. The thickness of the silicone film was measured under the stereomicroscope, and the gap between the groups was compared. Data was analysed using SPSS 22. Results: In group A, mean marginal gap was 82.43±25.00µm, and mean overall gap was 85.45±33.75µm. In group B, the corresponding values were 65.09±11.69µm and 78.04±26.67µm. There was a significant difference in the adaptation between the groups at the marginal and overall measurement points (p<0.05). Conclusion: Marginal and internal adaptations of polyether ether ketone coping crown for abutment could be considered clinically acceptable.
Publication Details
Page(s) 282-286
Published Journal: Journal of Pakistan Medical Association, Volume: 74, Issue: 2, Year: 2024

Request the full text from PASTIC

Keywords
Dental Implants, Crowns, Dental marginal adaptation, Dental abutment, Polyetheretherketone.