Comparison of point-of-care hemoglobin A1C testing with central laboratory method for its use in screening and monitoring of diabetes mellitus in a resource-constrained setting

Author(s)
Sara Reza Department, Quaid-e-Azam Medical College,Bahawalpur,Pakistan
Hafiz Muhammad Salman Yousaf Department, Quaid-e-Azam Medical College,Bahawalpur,Pakistan
Lubna Sarfraz Department, Quaid-e-Azam Medical College,Bahawalpur,Pakistan
Adil Mehmood Bahawal Victoria Hospital, Bahawalpur Pakistan
Abstract
Objective: The aim of this study is to evaluate the analytical performance of the Point-Of-Care Testing HbA1c analyzer for its use as an alternate to the central/core laboratory method. Material and Methods: This was a cross-sectional study in which 176 samples of HbA1c were analyzed by POCT HbA1c analyzer (A1C EZ 2.0) and central laboratory analyzer (Cobas 6000) as comparative instrument. Results: Based on the Bland-Altman plot, the HbA1c level for 164 out of 176 (93.18%) subjects analyzed by the POCT-HbA1c was within the range of the 95% limit of agreement. The area under the receiver operating characteristic (ROC) curve for clinical diagnosis of diabetes was 0.91. The sensitivity and specificity for HbA1c POCT were 94.5% and 88.0%. Using regression analysis, the slope of the A1C EZ 2.0 vs the Cobas 6000 Analyzer was 0.934, with an intercept of 0.20 and a correlation coefficient of 0.97. Conclusion: This study concludes that the POCT-HbA1c analyzer is a suitable replacement for central/core laboratory HbA1c measurement. It is useful for screening and monitoring of diabetes mellitus in certain situations especially in a resource constraint setting.
Publication Details
Page(s) 68-73
Published Journal: Pakistan Journal of Pathology, Volume: 34, Issue: 3, Year: 2023

Request the full text from PASTIC

Keywords
Diabetes Mellitus, Diabetes Mellitus, screening, HbA1C, monitoring, Point-of-care testing