Objective: To determine the oxygen reserve index (ORI) as a supporting parameter to the arterial partial oxygen pressure (Pa O2) in blood gases in hypoxia and hyperoxia monitoring with di erent fresh gas ows (FGF) in patients undergoing abdominal surgery. Study Design: Randomised controlled trial. Place and Duration of the Study: Department of Anaesthesiology and Reanimation, Samsun Education and Research Hospital, Turkey, from January to September 2020. Methodology: The study population of ninety patients was divided into three groups. After the high- ow period, the inspired oxygen fraction (FiO2) and ow-guided ventilation (FGF) were set to be 4 L/m and 40% in Group H (high- ow), 1 L/m and 50% in Group L (lowow), and 0.5 L/m and 68% in Group M (minimal- ow), respectively. Results: There was a very high statistically positive correlation between PaO2 and ORI in H, L, and M groups. When using a cut-o value of 0.005 for ORI for the detection of PaO2 >100 mmHg, the area under the curve (AUC) was 0.97 (p150 mmHg in the estimation of hyperoxia. Conclusion: ORI can be used to complement S2pinO low- ow anaesthesia in patients undergoing abdominal surgeries, provide guidance for PaO2, give information about tissue oxygen delivery, and contribute to the individualisation of oxygen therapy, and will therefore be included in the standard monitoring in the future.