Objective: To assess the variation in intraocular pressure measurements between Ocular Response Analyzer (ORA) and Goldmann Applanation Tonometer (GAT) in myopic patients undergoing laser assisted in-situ keratomileusis (LASIK) and photorefractive keratectomy (PRK). Study Design: Quasi-experimental study. Place and Duration of the Study: Armed Forces Institute of Ophthalmology (AFIO), National University of Medical Sciences, Rawalpindi, Pakistan, between September 2020 and 2021. Methodology: Myopic patients undergoing LASIK and PRK during the study period were selected. Baseline examinations and postoperative follow-ups were carried out to measure intraocular pressure at 1, 3, and 6 months after LASIK or PRK, using Goldmann Applanation Tonometer (GAT) and Ocular Response Analyzer, corneal compensated IOP (ORA IOPcc). Results: One hundred and thirteen eyes underwent myopic refractive surgery, LASIK (n = 60) or PRK (n = 53). Mean age of patients was 23.6 ± 4.11 years in the PRK group and 24.4 ± 5.94 years in LASIK group. For the PKR group, the GAT IOP value increased at 1-month (p <0.001), decreased at 3-month (p <0.001) and further decreased at 6-month (p <0.001) follow-up postoperatively. In the LASIK group, the GAT IOP value decreased at all time points (p <0.001). In both groups, the mean ORA IOPcc value increased at 1 month (p <0.001), decreased at 3-month (p <0.001) and further decreased at 6-month (p <0.001) follow-up postoperatively. The IOP mean values were generally found to be higher when measured using ORA as compared with GAT. Conclusion: Corneal refractive surgery markedly decreased IOP. This decrease in IOP was observed more after LASIK than after PRK. ORA was less likely to underestimate the intraocular pressure than GAT.