To compare Air puff tonometer to Goldman Applanation tonometer in different corneal thickness measurements and to find out the corneal thickness for which Air Puff measurements are closely related to Goldman tonometer.Cross sectional comparative study. Ophthalmology department Madina Teaching hospital Faisalabad from February 2016 to August 2016.Study was comparative cross sectional, conducted in outpatient department of Madina Teaching Hospital Faisalabad, Pakistan. Total of 50 subjects i.e. 100 eyes were checked by random sampling procedure considering inclusion criterion, for intraocular pressure by both Air puff and Goldman Applanation tonometer, corneal thickness was checked with automated corneal topographer. In younger subjects of age 15-28years old patients were found of corneal thickness is of average category of value 478-533nm, Adult was assigned the name of subjects of age 29-42years and corneal thickness among this age was equal in number in moderate 478-533nm, and thick cornea 534-589nmgroups. Old was assign name of group of subjects among age 42-55years and corneal thickness among this age group was maximum in thick cornea group. IOP by Air Puff tonometer measured in thin cornea (422-477nm) was 16.88+-2.78, in moderate corneal thickness (478-533nm) was 15.99+_2.93 and in thick cornea(534-589nm) was 17.15+_3.09.while IOP measured by Goldmann in thin corneas 15.38-+_2.93 for moderate corneal 14.97+-3.18 for thick corneas 16.56+_2.83.There is insignificant relationship between age and central corneal thickness. Air puff overestimates intraocular pressure than Goldmann tonometer. Goldmann Applanation tonometer and air puff tonometer are comparable with minimum variation in moderate corneal thickness(478-