The absolute quantification of total desmosine (DES/IDS) in human sputum was determined utilizing an LC-MS/MS technique employing synthesized internal standard. The procedure includes tandem mass spectrometric analysis (LCMS/MS), solid phase extraction (SPE), liquid chromatographic separation, and acid hydrolysis. From 0.05 (LLQ) to 20.0ng/mL, the calibration curve displayed acceptable precision and accuracy. The assay's precision and accuracy within and between days were calculated. Precision inter-assay ranged from 2.72 to 20.15%, whereas precision intraassay ranged from 4.76 to 23.47%. While the inter-assay ranged from -13.33 to 3.33% and the intra-assay ranged from 20.0 to -20.0% according to the accuracy. All of the sputum samples' concentrations fall within the assay's reportable range, which was determined using 40 sputum samples as the reference range for total DES. These findings show that the established method was able to quantitatively measure total desmosine in human sputum samples in a sensitive, repeatable, and accurate manner. In addition, Desmosine has the potential to be a biomarker for elastin degradation in respiratory diseases.