No Detail Found!

Flow injection determination of Mo(VI) in freshwaters with luminol chemiluminescence detection.

Author(s)
Attiq-Ur-Rehman Department of Chemistry, University of Balochistan, Quetta, Pakistan
Mohammad Yaqoob Department of Chemistry, University of Balochistan, Quetta, Pakistan
Amir Waseem Department of Chemistry, University of Balochistan, Quetta, Pakistan
Abdul-Nabi Department of Chemistry, University of Balochistan, Quetta, Pakistan
Abstract
A flow injection method is reported for the determination of Mo(VI) based on luminol chemiluminescence. The molybdophosphoric heteropoly acid formed by phosphate and ammonium molybdate in acidic conditions generated chemiluminescence emission due to the oxidation of luminol. The limit of detection was 0.5 ng mL-1 with a sample throughput of 120 h-1. A linear calibration graph was obtained over the range 1.0-200 ng mL-1 (r2 = 0.909; n = 9) with relative standard deviation (n = 4) in the range of 1.0-2.5 %. Interfering cations were removed by passing the sample through an in-line iminod iacetate-chel ating column. Analysis of a coastal water certified reference material (CASS-3) for Mo(VI) gave good results (8.6 ± 0.47 ng mL-1 compared with certified value of 8.95 ± 0.26 ng mL-1). The method was applied to freshwater samples and the recfoveries of 95-105 % Mo(VI) were observed.
Publication Details
Page(s) 74-79
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 31, Issue: 1, Year: 2009

Request the full text from PASTIC

0

Citations

0

Downloads

62

Views