No Detail Found!

Flow injection spectrophotometric determination of total iron in fresh waters using 1,10-phenanthroline reagent.

Author(s)
Attiq-Ur-Rehman Department of Chemistry, University of Balochistan, Quetta, Pakistan
Amir Waseem Department of Chemistry, University of Balochistan, Quetta, Pakistan
Mohammad Yaqoob Department of Chemistry, University of Balochistan, Quetta, Pakistan
Abdul-Nabi Department of Chemistry, University of Balochistan, Quetta, Pakistan
Abstract
A simple flow injection method is reported for the determination of total iron in fresh waters based on spectrophotometric detection of reddish orange iron (ll)-1,10-phenanthroline complex measured at wavelength 510 nm. The system permits high throughput of 120 samples per hour. A linear calibration (r2 = 0.9981) graph was obtained over the concentration range 0.2-1.0 mg L-1 with relative standard deviation (n = 4) 0.6-1.0 %. The limit of detection (3σ blank) was 0.01 mg L-1. The proposed method has been applied successfully for the determination of iron in fresh water samples and the results obtained (0.04 ± 0.003-0.09 ± 0.01 mg L-1) were in reasonable agreement with the results obtained using a spectrophotometric method (0.05±0.008-0.1±0.005 mg L-1) as the reference method
Publication Details
Page(s) 836-839
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 30, Issue: 6, Year: 2008

Request the full text from PASTIC

0

Citations

1

Downloads

70

Views