Comparison of maximal infusion rate of FMS 2000 fluid warmer on various catheters and fluids.

Author(s)
Yong Han Kim Departments of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Young Gyun Choi Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Sira Bang Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Sang Eun Lee Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Jae Hong Park Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Sang Yoon Jeon Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Jung Dae Shin Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Ki Hoon Kim Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Jin Soo Kim Department of Anesthesiology and Surgery, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea
Abstract
Objective: To compare the maximal infusion rate of FMS 2000 fluid warmer on various intravenous catheters and fluids. Methods: This study was performed on 4th February 2011 in operation room, Haeundae Paik Hospital, Inje University, Busan, Republic of Korea. 0.9% sodium chloride (NS), Ringer’s lactate (RL), and 6% hydroxylethyl starch 130/0.4 (HES) were applied to FMS 2000 fluid warmer. After FMS 2000 was equipped with one kind of fluid, 16 types of catheter including double lumen catheter were connected to the end of line. We accelerated infusion rate to 750ml/min under 300mmHg of pressure limit. Maximal infusion rate under 300mmHg was measured three times. Maximum value of maximal infusion rate was compared. Results: 7-, 8.5F rapid infusion catheter (RIC) resulted in rate of 750ml/min in NS, RL and HES group. Other catheters did not exceed 750ml/min. Maximal infusion rate of NS and RL was faster than HES in all kinds of catheter except RIC. There was no statistically significant difference between NS and RL. Conclusion: Maximal infusion rate of HES is slower than NS and RL in ordinary sized venous catheter using FMS 2000. NS and RL had no difference in maximal infusion rate.