Accuracy of working length measurement with endo motor having built-in apex locator and comparison with periapical radiographs

Author(s)
Syed Yawar Ali Abidi Sindh Institute of Oral Health Sciences, Jinnah Sindh Medical University, Karachi, Pakistan
Marium Azfar Sindh Institute of Oral Health Sciences, Jinnah Sindh Medical University, Karachi, Pakistan
Talha Nayab Sindh Institute of Oral Health Sciences, Jinnah Sindh Medical University, Karachi, Pakistan
Asjad Shaukat Ishrat-Ul-Ebad Institute of Oral Health Sciences, Dow University, Karachi, Pakistan
Mehvesh Hasan Ishrat-Ul-Ebad Institute of Oral Health Sciences, Dow University, Karachi, Pakistan
Nabeel Naeem Baig College of Physicians and Surgeons Pakistan.
Khadijah Abid College of Physicians and Surgeons Pakistan.
Abstract
Objective: To determine the accuracy of working length measurement with endomotor having builtin apex locator by comparing it with periapical radiographs. Methods: The cross-sectional study was conducted at the Department of Operative Dentistry, Dr Ishrat-ul-Ibad Khan Institute of Oral Health Sciences, Dow University of Health Sciences, Karachi, from January to June 2014, and comprised permanent maxillary anterior teeth with mature apices. A preoperative radiograph was taken by using standardised paralleling technique. The access cavity was prepared by a high speed water-cooled diamond bur. After doing initial filing and drying the canal, the working length was taken with an endomotor with built-in apex locator in dual mode using the Protaper Rotary File S1. The length was measured using a millimetre scale. After attaining tentative working length from the pre-operative radiograph, 1mm was subtracted from the radiographic apex. A stainless steel K-file was placed in the canal at this length and a radiograph was taken by paralleling technique using intraoral film holder. The electronic and radiographic findings were noted. Accuracy was considered positive if the endomotor reading was 0-2mm short of the radiographic working length. Data was analysed using SPSS 16. Results: Of the 96 cases, working length with X-Smart Dual was acceptable in 85(88.5%) cases, and unacceptable in 11(11.5%) cases. Conclusion: Accuracy of working length measurement with endomotor having built-in apex locator was found to be a better measuring tool compared to the conventional periapical radiographs.
Publication Details
Page(s) 437-441
DOI DOI not available
Published Journal: Journal of Pakistan Medical Association, Volume: 70, Issue: 3, Year: 2020

Request the full text from PASTIC

Keywords
X-Smart Dual, Working length, Radiographs.