No Detail Found!

Positional syncope occurring just at the right time in a patient with a permanent pacemaker: A case report

Author(s)
Syed Haseeb Raza Naqvi National Institute of Cardiovascular Diseases,Karachi,Pakistan
Madiha Fatima National Institute of Cardiovascular Diseases,Karachi,Pakistan
Pir Sheeraz Ali National Institute of Cardiovascular Diseases,Karachi,Pakistan
Maqsood Alam National Institute of Cardiovascular Diseases,Karachi,Pakistan
Muhammad Taha Khan National Institute of Cardiovascular Diseases,Karachi,Pakistan
Faisal Qadir National Institute of Cardiovascular Diseases,Karachi,Pakistan
Abstract
We present an unusual case of positional syncope occurring years after pacemaker implantation. Syncope was caused by a pacemaker lead fracture, which resulted from subclavian crush syndrome. The syncope occurred incidentally during hospital admission and was timely diagnosed using an integrated approach of history taking, examination if ndings, device interrogation and radiographic parameters. The patient subsequently underwent lead and device alteration which led to resolution of her symptoms.
Publication Details
Page(s) 2461-2463
DOI DOI not available
Published Journal: Journal of Pakistan Medical Association, Volume: 71, Issue: 10, Year: 2021

Request the full text from PASTIC

Keywords
pacemaker, syncope, subclavian crush, lead fracture