Usefulness of forehead infrared thermometers to screen patients for fever during covid-19 pandemic

Author(s)
Nauman Sarir Khan Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Abdul Rehman Arshad Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Muhammad Tariq Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Maqsood Khan Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Muhammad Bilal Siddique Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Khawar Shabbir Combined Military Hospital, Peshawar/National University of Medical Sciences (NUMS), Pakistan
Abstract
Objective: To determine accuracy of infrared thermometer for detection of fever as compared to mercury thermometer. Study Design: Cross sectional study. Place and Duration of Study: Department of Medicine, Combined Military Hospital Peshawar, from May to Jun 2020. Methodology: All willing adult patients reporting to the fever desk were selected by consecutive sampling. Exclusion criteria included any dermatological condition affecting forehead and unwillingness. Forehead temperature was first checked twice using Kinlee FT3010 infrared thermometer. Axillary temperature was then recorded using a standard clinical mercury thermometer. Results: There were 538 patients, including 251 (46.65%) males and 287 (53.35%) females, aged 46.76 ± 12.44 years. Median temperatures recorded with infrared and mercury thermometers were 97.00°F (interquartile range: 95.10- 97.80°F) and 98.30°F (interquartile range: 98.00- 98.90°F) respectively (p<0.001). Intra-class correlation was 0.143 (95% CI -0.052, 0.323). There was a weak to moderate correlation (R: 0.366; p<0.001) between temperatures recorded by the two techniques. ROC curve analysis for temperatures recorded by infrared thermometer revealed an area under curve of 0.725 at a threshold of 98.6°F and 0.746 at a threshold of 100.4°F defined by mercury thermometer. Infrared thermometer had sensitivity, specificity, positive predictive value and negative predictive value of 13.61% and 9.38%, 97.95% and 99.80%, 71.43% and 75.00%, and 75.10% and 95.57% for thresholds of 98.6°F and 100.4°F respectively. Conclusion: Infrared thermometer underestimates temperatures recorded by mercury thermometer. Limits of agreement are too broad, indicating inconsistency in measurements. A significantly lower threshold is required to improve the sensitivity of Infrared thermometer in picking up fever.
Publication Details
Page(s) 597-602
DOI DOI not available
Published Journal: Pakistan Armed Forces Medical Journal, Volume: 70, Issue: 2, Year: 2020

Request the full text from PASTIC

Keywords
agreement, Temperature measurement, Skin temperature, infrared thermometry, Thermometry