Study Objective: To compare diagnostic accuracy o f predicted clearance method, Gates method, CockcroftGault method and plasma 1- sample clearance method with plasma 2-samples clearance method with Tc99m DTPA for the estimation o f glomerular filtration rate(GFR). Study Design: Comparative study. Materials and Methods: This study included 91 consecutive patients who were referred for evaluation of renal function to the Nuclear M edicine section o f Karachi Institute o f Radiotherapy and Nuclear Medicine (KIRAN) from September 2004 to September 2005. The GFR was determined simultaneously by 5 methods including Plasma two-Sample Clearance method after Tc-99m DTPA injection (PSC 2); Plasma one-Sample Clearance m ethod after Tc-99 m DTPA injection (PSC 1); Gamma camera uptake m ethod after Tc-99m DTPA injection (Gates method); Predicted Creatinine Clearance by Modification of Diet and Renal Diseases (MDRD); and Cockcroft-G ault's equation for GFR estimation (CG). PSC 2 was chosen as a reference. Results: Out o f the 91 patients, 71 were males and 20 females with age ranging from 16-68 years. The regression equation of the PSC 1, Gates, MDRD and CG method against the PSC 2 was Y = 1.884 +0.970X (r=0.90, p<0.001, SEE value=10.23 ml/min/1.73m2), Y = - 9.944 + 1.083X (r=0.82, p<0.001, SEE value=11.02 ml/min/1.73m2), Y =25.606+ 0.640X (r=0.71, p=0.002, SEE value=15.56 ml/min/1.73m2), and Y =14.981+ 0.714X (r=0.77, p=0.002, SEE value=14.44 ml/min/1.73m2) respectively. In comparison with the GFR by PSC 2, the PSC 1 and Gates tended to overestimate by 1% (p=0.359) and 2% (p=0.265) respectively, MDRD and CG tended to underestimate GFR by 11% and 14% respectively (p<0.001). Conclusions: PSC 1 correlate well w ith PSC 2 and either can be substituted for the other as ideal GFR markers. The Gates method shows good correlation with PSC 2 however it is less precise than PSC 1. MDRD and CG m ethods due to sig n ifican t u n d erestim atio n are not co n sid ered as id eal G FR m arker.