Comparison of Axial Ocular Measurements with Contact and Non-Contact Biometry

Author(s)
Muhammad Arslan Ashraf Department of Ophthalmology KEMU, Lahore, Pakistan
Muhammad Suhail Sarwar Department of Ophthalmology KEMU, Lahore, Pakistan
Muhammad Awais Afzal Lahore General Hospital Lahore, Pakistan
Imran Khalid Services Hospital, Lahore, Pakistan
Sehrish Shahid Department of Ophthalmology KEMU, Lahore, Pakistan
Abstract
Purpose: To compare between non-contact and contact biometry for measurements of central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT) and axial length (AL). Material and Methods: Eighty-four subjects, (168 non-pathological eyes) visiting the eye outpatient's department were recruited by non-probability convenience sampling. Patients with high refractive errors and suffering from any ocular pathology were excluded from the study. CCT, ACD, LT and AL were measured with non-contact Biometer (HAAG Streit) followed by Contact Biometer (Ultrasound) after taking consent from the patient. Data was entered and analyzed by using SPSS-21 and Medcalc software. Normality of quantitative data was checked with Shapiro Wilk test. Independent sample t test was used for parametric variable and Mann Whitney-U test was used for non-parametric data. For the agreement between two techniques Cohen's Kappa test used and Bland-Altman plot was drawn for graphical presentation. p-value equal or less than 0.05 was taken as significant. Results: Mean age of 84 subject (female: 45.24% and male: 54.76%) was 53.05 ± 13.56 years. The AL was significantly longer for the non-contact measurement with the difference of 0.53 ± 0.32 mm (p < 0.001). Contact pachymetry was significantly higher with the difference of 8.67 ± 20.83 µm (p = 0.046). ACD was significantly deeper for non-contact measurements with the difference of 0.51 ± 0.32 mm (p < 0.001). Contact ultrasound Ascan measured LT significantly thicker with the difference of 0.59 ± 0.56 mm (p < 0.001). Conclusion: There is significant difference of axial ocular measurements (CCT, ACD, LT and AL) between contact (ultrasound A-scan) and non-contact (HAAG Streit) biometry (p < 0.05).
Publication Details
Page(s) 72-78
DOI DOI not available
Published Journal: Pakistan Journal of Ophthalmology , Volume: 36, Issue: 1, Year: 2020

Request the full text from PASTIC

0

Citations

0

Downloads

55

Views