No Detail Found!

Determination of Related Substances in Ketoprofen Injection by RP-HPLC Method.

Author(s)
Lei LUO College of Pharmaceutical Sciences, Southwest University, Chongqing, China
Mingguo Tan College of Pharmaceutical Sciences, Southwest University, Chongqing, China
Yonghuang LUO College of Pharmaceutical Sciences, Southwest University, Chongqing, China
Abstract
The paper aims to establish a RP-HPLC method for the simultaneous determination of six related substances in ketoprofen injection. The separation was performed on a VP-ODS C18 column (4.6mm×250mm, 5µm) with the mobile phase of 6.8% phosphate buffer solution (adjusted to pH3.5 with 85% phosphoric acid)-acetonitrile-water (2:43:55,v/v/v) at a flow rate of 1.2mL·min-1. The detection wavelength and the injection volume were set at 233nm and 20µL, respectively. Impurity A and C were calculated by external standard method. Main component self-compare method with calibration factor was used to calculate impurity B, D, E, F and main component self-compare method without calibration factor was used to calculate unspecified impurity. Related substances and degraded substances were completely separated from ketoprofen. For impurity A and C, the linear range of determination were separately 0.06 µg·mL-1 ~ 3.6µg·mL-1 and 0.036µg·mL-1 ~ 2.4µg·mL-1 with the correlation coefficient of 0.9999. The average recoveries (n=9) were 98.13% (RSD=0.35%) and 96.32% (RSD=0.43%). The precision and repeatability for method were good. With reference to ketoprofen (retention time =10.06 min), the relative retention time of impurity B, D, E, F were 0.71, 1.46, 0.59, 2.13, respectively, and the relative correction factors were 0.962, 0.938, 0.957, 0.960, respectively. Finally, determined that the contents of impurity A could not be more than 0.3%, any of the contents of impurity B, C, D, E, F and unspecified impurities could not be more than 0.2%, sum of the contents of impurities other than A and C couldn't be more than 0.5%. The method was proved to be simple, rapid, accurate, sensitive and suitable for the simultaneous determination of six related substances in ketoprofen injection.
Publication Details
Page(s) 1607-1614
DOI DOI not available
Published Journal: Pakistan Journal of Pharmaceutical Sciences, Volume: 32, Issue: 4, Year: 2019

Request the full text from PASTIC

0

Citations

0

Downloads

82

Views