This paper describes the preparation as well as evaluation of biocompatible membranes and coatings, for an amperometric glucose sensor in order to extend the linear range of the sensor characteristics. For this purpose, both Nafion and polyurethanes are highly suitable, as they increase the upper limit of linearity up to a factor of five. Nafion membranes underwent optimization of their diffusion behavior by temperature-controlled pressing. Sensors with coatings showed much smaller response times (one minute) than with membranes. In both cases, we achieved detection limits of 0.3 mmol/l glucose with a linear sensor characteristic of up to10 mmol/l compared to1.4 mmol/l for the unmodified electrodes. Furthermore, polymer coatings and membranes prevent clogging of the sensors by the bioanalytes.
Publication Details
Page(s) 17-22
DOI DOI not available
Published Journal: Journal of Chemical Society of Pakistan, Volume: 35, Issue: 1, Year: 2013

Request the full text from PASTIC

0

Citations

0

Downloads

204