Abstract
Background: Endotracheal cuf, when inflated to extremely high pressures lead to tracheal ischemia due to reduced tracheal blood supply while faulty ventilation and aspiration of gastric content or accidental extubation can occur when underinflated. In this study we evaluate the effects of maintaining fixed optimal cuf pressure (25 cmH2O) during endoscopic anterior cervical spine surgeries, in terms of postoperative airway complications. Methodology: In our study 100 cases were scheduled to undergo elective anterior cervical spine surgeries under general anesthesia (GA). These cases were divided into two groups according to cuf inflation technique. In Group P (study group), endotracheal tube cuf (ETTc) was inflated with air syringe by senior anesthesiologist. Cuf pressure was estimated by feel of pilot balloon and by absence of audible air leak. Later endotracheal tube cuf pressure (PETTc) was recorded and monitored every 30 min by Portex™ Cuf Inflator pressure gauge. In Group C (control group), ETTc was inflated by attaching with Portex™ Cuf Inflator pressure gauge and pressure was sustained at 25 cmH2O till patient remains intubated. Both groups were observed for development of any airway complications at 24 hours post-extubation. Results: There was insignificant diferences in age, sex, height, weight, basal metabolic index (BMI), and end tidal CO2 (EtCO2), surgery duration and duration of endotracheal intubation in both groups. Mean measured PETTc was 31.25 ± 6.78 mmHg after fixing the EET in Group P. The frequency of postprocedural cough and sore throat in Group P was significantly higher than in the control Group. Data was expressed in Mean ± SD. A statistically significant rise in PETTc in Group P was found at different interval of time. Conclusion: We concluded that maintaining optimal cuf pressure with cuf inflator pressure gauge is effective in minimizing postoperative common airway complications. Additionally use of personal judgement of estimating PETTc is not reliable and hence should be discouraged.
Publication Details
Page(s) 442-446
DOI DOI not available
Published Journal: Anaesthesia, Pain and Intensive Care, Volume: 21, Issue: 4, Year: 2017

Request the full text from PASTIC

Keywords
Surgery, Cervical Spine, Airway, Endotracheal tube, Endotracheal tube, Endotracheal tube, Endotracheal tube, cuff pressure, Postoperative complications
References
DullenkopfA,GerberA,WeissM., .Fluid leakage past tracheal tube cuffs: Evaluation of the new Microcuff endotracheal tube,Intensive Care Med 29 1849 -53
LienTC,WangJH, .Incidence of pulmonary aspiration with different kinds of artificial airways,Zhonghua Yi Xue Za Zhi (Taipei) 49 348 -53
TrivediL,JhaP,BajiyaNR,TripathiD., 2010.We should care more about intracuff pressure: the actual situation in government sector teaching hospital,Indian J Anaesth 10 0019 -5049
FanCM,KoPC,TsaiKC,ChiangWC,ChangYC,ChenWJ,YuanA, .Tracheal rupture complicating emergency endotracheal intubation,Am J Emerg Med 22 289 -93
JainMK,TripathiCB, 2011.Endotracheal tube cuff pressure monitoring during neurosurgery-Manual vs Automatic method,J Anesthesiol Clin Pharmacol 10 0970 -9185
NseirS,BrissonH,MarquetteCH,ChaudP,Di PompeoC,DiarraM, 2009.Durocher A. Variations in endotracheal cuff pressure in intubated critically ill patients: prevalence and risk factors,Eur J Anaesthesiol 26 229 -34
HoffmanRJ,ParwaniV,HahnIH, .Experienced emergency medicine physicians cannot safely inflate or estimate endotracheal tube cuff pressure using standard techniques,Am J Emerg Med 24 139 -43
LiuJ,ZhangX,GongW,LiS,WangF,FuS, 2010.Correlations between controlled endotracheal tube cuff pressure and postprocedural cmplications: A multicenter study,Anesth Analg 111 1133 -7
AbdelattiMO, .A cuff pressure controller for tracheal tubes and laryngeal mask airways,Anaesthesia 54 981 -986
SeegobinRD,Van HassletGL, .Endotracheal cuff pressure and tracheal mucosa blood flow: endoscopic study of effects of four large volume cuffs,1984 March 288 965 -8
RaederJC,BorchgrevinckPC,SellevoldOM, .Tracheal tube cuff pressures. The effect of different gas mixtures,Anaesthesia 40 444 -7
CombesX,SchauviliegeF,PeyrousetO,MotamedC,KirovK,DhonneurG, .Intracuff pressure and tracheal morbidity: influence of filling cuff with saline during nitrous oxide anesthesia,Anesthesiology 95 1120 -4
SoleML,PenoyerDA,SuX,JimenezE,KalitaSJ,PoalilloE, 2009.Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study,Am J Crit Care 10 133 -43
.Guidelines for the management of adults with hospital-acquired, ventilator-associated and healthcareassociated pneumonia,2005 Feb 15 388 -416
JanossyKM,PullenJ,YoungD,BellG., 2010.The effect of pilot balloon design on estimation of safe tracheal tube cuff pressure,Anaesthesia 06413 2044 -791
WyneJM,Jones KL, .Tongue cyanosis after laryngeal airway insertion,Anesthesiology 80 1403 -
StanleyTH, .Nitrous oxide and pressures and volumes of high and low pressure endotracheal tube cuffs in intubated patients,Anesthesiology 42 637 -40
MandoeH,NikolajsenL,LintrupU,JepsenD,MolgaardJ, 1992.Sore throat after endotracheal intubation,Anesth Analg 74 897 -900
Curiel GarciaJA,Guerrero-RomeroF,Rodriguez-MoranM, 2001.Cuff pressure in endotracheal intubation: should it be routinely measured? Gac Med Mex, 137 179 -82
SvensonJE,LindsayMB,O'ConnorJE, .Endotracheal intracuff pressures in the ED and prehospital setting: is there a problem?,Am J Emerg Med 25 53 -6
BlanchPB, 2004.Laboratory evaluation of 4 brands of endotracheal tube cuff inflator,Respir Care 49 166 -173
WyneJM,Jones KL, .Tongue cyanosis after laryngeal airway insertion,Anesthesiology 80 1403 -
NagaiK,SakuranmotoC, .Unilateral hypoglossal nerve paralysis following the use of the laryngeal mask airway,Anaesthesia 49 603 -4
BrandtL,ParkerH., .The rediffusion system: limitation of nitrous oxideinduced increase of the pressure of endotracheal tube cuffs,Anaesthesist 32 459 -64
McHardyFE,ChungF, .Postoperative sore throat: cause, prevention and treatment,Anaesthesia 54 444 -53
SuzukiN,KooguchiK,MizobeT,HiroseM,TakanoY,TanakaY., .Postoperative hoarseness and sore throat after tracheal intubation: effect of a low intracuff pressure of endotracheal tube and the usefulness of cuff pressure indicator,Masui 48 1091 -5
OzerAB,DemirelI,GunduzG,Erhan OL, 2013.Effects of user experience and method in the inflation of endotracheal tube pilot balloon on cuff pressure,Niger J Clin Pract 10 110139 -3077
TuHN,SaidiN,LeiutaudT,BensaidS,MenivalV,DuvaldestinP, 1999.Nitrous oxide increases endotracheal cuff pressure and the incidences of tracheal lesions in anesthetized patients,Anesth Analg 89 187 -90
PriebeHJ, 2000.N2O and endotracheal cuff pressure,Anesth Analg 90 230 -231
0

Citations

0

Downloads

66

Views